Enhanced phase identification with simultaneous
STEM EDS and EELS/ELNES
In problems involving real-world materials, the key to a solution often lies in using multiple complementary (and/or redundant) sources of information to discern and identify the various phases that might form during material production and processing. Of particular interest in this respect are unexpected and possibly undesired reaction products that can develop at interfaces between different components of an engineered structure. In such cases, it can be quite beneficial to simultaneously monitor both elemental composition (EDS fluorescence peaks and EELS core-edge signals) and electronic structure (energy-loss near edge structure or ELNES) signals to form a more complete picture of the interfacial phases present.
Gatan offers an array of products to enable this sort of concurrent multimode data acquisition on your analytical TEM/STEM system. Start with STEMPack, which provides multi-signal digital STEM imaging and spectrum-imaging. Add the EDS Acquisition option of our GMS (Gatan Microscopy Suite) software to systematically capture fluorescent X-ray spectra from your existing EDS system. Then complete the system with either a GIF Tridiem or Enfina electron energy-loss spectrometer to start acquiring simultaneous EDS, EELS, and ELNES data sets from your materials.
Read on for a detailed look at how STEMPack with EDS option and GIF Tridiem were applied to the solution of a coating adhesion problem in a very practical and commercially important steel pipe material.
A case study: identification of interfacial phases in coated steel via simultaneous STEM EDS, EELS, and ELNES with STEMPack/EDS and GIF Tridiem
The sample investigated was a section of steel pipe with a multi-layer wear- and corrosion-resistant coating developed by Dr. Deepak Upadhyaya of Sub-One Technology (www.sub-one.com). An important physical parameter of this coating is its adhesion to the underlying steel. To achieve an understanding of the coating adhesion it was important to learn the nature of the phases that form at the steel-coating interface.




In Z-contrast mode, specimen areas of higher average atomic number (Z) appear brighter than those of lower average Z. The 20 nm interfacial layer is now readily apparent and clearly has a lower average Z than the underlying steel (since it has a darker contrast level than the steel). This HAADF STEM image then became the reference point for setup of a systematic STEM spectrum image (STEM SI) acquisition with simultaneous EELS and EDS spectra.

Click here to download this movie in AVI format
Click here to download this movie in WMV format
There are several advantages to systematically collecting large numbers of spectra in such spectrum image data sets, as opposed to a few brief point analyses. One major advantage is the unbiased and redundant nature of the sampling, allowing immediate and direct confirmation that spectral features are real and reproducible, not spurious.
Another major benefit is the fact that both spatial and spectral information is permanently recorded. This allows one to effectively play back the experiment in subsequent off-line data processing sessions free of the constraints of the TEM lab. In this way, the results may be examined in greater detail to reveal previously unexpected and unnoticed specimen components. The concomitant capture of spatial and spectral information also makes subsequent correlation of elemental composition with specific sample areas entirely unambiguous and robust, particularly in the extraction and precise spatial registration of elemental composition maps.
Finally, another key advantage of spectrum image acquisition is the distribution of dose, both with respect to specimen damage and the accumulation of carbon contamination, a common problem in STEM-based point analysis. One can thus harvest artifact-free spectra with high total counts (and low statistical noise) by summing spectra over selected homogeneous sample regions (as done in the above line-scan movie across the interface).
Viewing the animated line scan above makes it very clear that the specimen composition goes through abrupt changes in the vicinity of the interface. As anticipated by the EFTEM and Z-contrast images, the 20 nm layer between the steel and overlying coating clearly has a distinct composition with relatively high oxygen content, as might be expected for a surface oxide layer on the steel. However, a closer examination of the line scan movie at slower frame rate reveals the presence of relatively high oxygen content in a thin layer immediately above the oxidized steel. Such observations made during off-line exploration of the SI data set help guide subsequent sub-sampling of the data set for in-depth quantitative analysis.



The complementary nature of EELS and EDS elemental signals and the value of having both available in STEM SI are clearly illustrated by the above spectra. In the EELS energy-loss range probed, accurate extraction and quantification of the Si L edge signal is problematic, but the Si K signal in EDS is easily captured and quantified. In the EDS data, oxygen quantification becomes difficult when transition metals Ti, V, Cr, or Mn are present, but the EELS results above show clear separation and reasonable quantification of the O and Cr signals. Due to the secondary and fluorescent nature of EDS signals, spurious Cu peaks from the support grid or TEM system components are common problem, as seen in the EDS data above. EELS measures primary events captured in the direct beam and is free of such artifact signals, as clearly demonstrated by the absence of a visible Cu signal in any of the above EELS spectra.
Taken together, the key finding of the above data sets is that ROI 2 is an oxide layer that does not involve any of the metals from the underlying steel pipe material. It is quite distinct from the obvious native oxide layer in ROI 3 on the surface of the steel and appears to be dominantly an oxide of silicon.
Further evidence in favor of this conclusion comes from a detailed look at the EELS data in the vicinity of the oxygen K edge, the so-called energy-loss near edge structure (ELNES). ELNES reflects the unoccupied density of states in the electronic band structure of a material. This varies with the electron configuration and bonding associated with a given atom type and is thus another probe that can be used to differentiate chemical phases in a sample. The inherent eV-level (or better) resolution of EELS spectra acquired with Gatan's state-of-the-art EELS spectrometers mounted to present-day analytical TEM systems, means that ELNES signals are immediately and simultaneously available in virtually any routine EELS acquisition. No additional effort is required to harvest this information. One need only take a close look at it once the data sets have been acquired.
The spectra below show the O K ELNES for the two distinct interfacial oxide layers proposed for ROI 2 and ROI 3:


Summary and conclusion
The case study reviewed above demonstrates that simultaneous capture of multiple analytical TEM signals enable one to form a robust and consistent picture of the construction and composition of important material structures at the nanometer scale. For the particular material studied here, the detection and identification of two clearly distinct interfacial oxide phases on either side of the steel-coating interface was a key finding about the nature of the adhesion between coating and steel. These results led Sub-One Technology, the manufacturer of the coated pipe, to make refinements to their process and thereby improve the adhesion of the coating by 150%, a significant positive result for a practical, real-world material.
More Information
To learn more about Gatan's products for analytical TEM to see whether they might help solve some of your real-world materials problems, please follow these links:
If you would like to learn more about the research and development effort behind GIF Tridiem, please refer to the following publications:
H. A. Brink, M.M.G. Barfels, R.P. Burgner, B.N. Edwards, Ultramicroscopy 96 (2003) 367.
G. Kothleitner and F. Hofer, Micron 34 (2003) 211.