646 Double Tilt
Detailed analysis of crystalline TEM specimens requires
the application of electron diffraction in addition
to EDX. In these cases, the 646 Double Tilt
Analytical Holder has the capability of up to +/-60°
main tilt and up to +/-30° secondary tilt as well
as excellent low background performance. The ToggleTilt
drive mechanism used for the second tilt is simple,
reliable, and eliminates the need for complicated
tilt drive endstops. For TEMs with small objective
lens polepiece gaps, holders are available with thinner
cross-sections to provide higher tilt capability.
Special double tilt holders are also available having
electrical terminals close to the tilt cradle in order
to perform electrical experiments on the specimen
while it is in the TEM.
Analytical TEM
To obtain good EDX spectra and accurate compositional
analyses from specimens in a TEM, a well-designed
specimen holder is essential. The Gatan range of Analytical
TEM Holders meets the requirements for good EDX analysis
without compromising TEM image resolution.
Low Background Design
The beryllium specimen cup and clamping ring of the
single and double tilt analytical holders minimize
the non-specimen x-ray signal, particularly that from
the specimen holder. Beryllium is chosen for its low
fluorescent yield and for the low detection efficiency
of its characteristic radiation. Also, because the
characteristic x-rays from beryllium have such a low
energy, they are not able to cause other materials
around the specimen to fluoresce. Furthermore, beryllium,
unlike carbon or graphite, possesses good mechanical
properties, is UHV compatible, and is easy to clean.
Other parts of the specimen holder tip that cannot
be made from beryllium are mostly made from high-strength
aluminum alloys in a further attempt to minimize the
x-ray signal from the holder and the surrounding microscope.
This intelligent choice of materials makes the Gatan
analytical holders particularly suitable for high
spatial resolution EDX by STEM or small-probe TEM.
Reduced Shadowing
Having minimized the unwanted signal from the specimen
holder, the specimen signal is further optimized by
removing those parts of the specimen cup that are
along the line of sight of the EDX detector. In the
final design, low-angle detectors require minimal
tilting of the holder to stop shadowing, while high-angle
detectors require no tilting.
Integral Faraday Cup
Accurate quantitative analysis requires a knowledge
of the electron current incident on the specimen.
Gatan analytical holders have a miniature Faraday
cup in the holder tip, close to the specimen cup.
An electrical socket is provided at the end of the
holder to connect the Faraday cup to a picoammeter.
An alignment tool, which fits between the holder and
the TEM stage, allows the holder to be positioned
quickly and easily to place the Faraday cup exactly
in the path of the electron beam.
|