| Cryo
electron microscopy requires a special environment for the
microscope
By Linda Melanson, Gatan, Inc.
Working
at cryo temperatures is more demanding than working at room
temperature so special consideration must be taken with regard
to designing the space that will house the electron microscope.
In a transmission electron microscope, the tiny specimen grid
is suspended between the objective lens pole pieces of the
TEM by means of a specialized low temperature specimen holder.
Although one might think that the specimen is well protected
from the environment outside of the robust electron optical
column, it is actually quite susceptible to many factors that
can degrade the overall resolution of the final projected
image. It is important to minimize or totally eliminate any
environmental issues that would adversely affect the operation
of the electron microscope and ancillary instrumentation used
in conjunction with the microscope. Three key factors are
vibration, climate control and stray electromagnetic fields.
Vibration can be transmitted to the cryo specimen
holder either mechanically through the microscope or acoustically
through the air. Mechanical vibrations can occur at low and
high frequency. Specimen holders for Cryo TEM come in several
forms: side entry, fixed top entry and holders that combine
attributes of both. Although fixed top entry cryo holders
are known to be extremely stable due to their design characteristics;
all cryo holders can suffer the ill effects of vibration resulting
in degradation of the image. Some side entry cryo holders
that incorporate external liquid nitrogen dewar can be susceptible
to high frequency vibration if the level of liquid nitrogen
within the dewar is not maintained a constant level. In addition,
if the cooling water that flows through the objective lens
is not properly dampened, it will also cause vibration to
be transmitted to the specimen. Low frequency vibrations can
originate from traffic on a near-by streets, geoseismic activity,
activity at a near-by construction site, or even foot traffic
near the microscope. Unfortunately, vibrations in the low
frequency range are the most difficult to eliminate. Vibration
that is transmitted through the air is generally caused by
acoustic noise that comes from, for example, noisy air handling
equipment that services the microscope room. These services
should be supplied to the microscope room in a very quiet
and controlled manner. Even small air currents in the vicinity
of the specimen holder can degrade resolution. No matter what
the source, vibrations can limit resolution by causing blurring
of the image in the horizontal plane, especially in the direction
perpendicular to x axis for specimen translation (i.e. perpendicular
to the long axis of the specimen rod of a side entry cryo
holder), or blurring of the image along the vertical axis.
Many high resolution electron microscopy laboratories go to
great lengths to eliminate all sources of vibrations that
might affect the performance of the microscope.
Ideally, the electron microscope should be
housed in a room that has stringent humidity control (preferably
15% R.H. or lower) and temperature control. This is especially
true if one is going to use side entry cryo holders. Depending
upon the design of the holder, the tip where the specimen
is located is very cold when in use (approximately -180 °C)
and it acts as a cryo sorption pump. If the humidity in the
room is very high, a significant amount of water vapor in
the air will condense (as frost) onto the surface of the shields
that protect the cryo grid during transfer to the microscope
airlock. This condensed water vapor is then carried into the
airlock mechanism, and subsequently into the optical column,
eventually degrading the vacuum as well as acting as a source
of contamination to the cryo specimen. If the humidity in
the microscope room is maintained to a very low level, this
effect is minimized. Once the cryo specimen is loaded into
the microscope, it is necessary to use devices to prevent
the buildup of contamination onto the surface of the specimen.
Contamination can be minimized by using anti-contamination
devices provided by the microscope manufacturer or by third-party
instrument designers who specialize in instrumentation for
cryo EM such as Gatan. In addition, cryo specimens are particularly
susceptible to drift from temperature gradients, and variations
in these gradients, that can originate from both inside and
outside of the microscope. Drift occurs in a direction parallel
to the long axis of the mechanism that holds the specimen
holder. Room air temperature variation caused by the cycling
of heating or cooling air entering the microscope room, or
the variations in the temperature of the cooling water used
to cool the lenses of the microscope, can all result in specimen
drift which will limit the final resolution of the image.
External AC and changing DC fields can affect
the high resolution performance of the microscope. The electron
optical column is composed of a series of electromagnetic
lenses that are used to focus and to control the position
of the electron beam that is emitted from the electron gun.
Although the microscope column is shielded, external fields
can still penetrate the structure of the column. This results
in movement and subsequent blurring of the projected image.
Although AC and changing DC fields degrade images, changing
DC fields are particularly troublesome and have a greater
effect on the ultimate performance of the microscope than
AC fields. This field can originate from the movement of a
nearby elevator, commuter rail trains and even movement of
metal chairs in the vicinity of the microscope. In addition,
even a large constant DC field, such as from the superconducting
magnets found in NMR instruments, can affect performance.
Microscope manufacturers have stringent guidelines
regarding acceptable limits for the microscope site, especially
when the microscope will be used for collecting high resolution
cryo data. Many high resolution cryo TEM laboratories may
also employ outside testing firms to ensure that their facility
fully meets the required specifications. Meeting or exceeding
the microscope installation requirements when designing a
cryo electron microscopy laboratory will help to insure success
in implementing this exciting technique.
|