The GIF QuantumERS™ imaging filter represents the new standard in EELS and EFTEM instrumentation.  No other commercial system provides the level of EELS and EFTEM performance afforded by the QuantumERS™ imaging filter.  With its extreme spectral resolution and stability, it is the system of choice for high-resolution monochromated sources.  The GIF QuantumERS™ imaging filter includes all the features and advantages of the GIF QuantumER™ imaging filter with the added feature of enhanced stability electronics, hand-tuned and matched components, and extensive tuning, testing and qualification of all key systems. 

The GIF QuantumERS™ imaging filter is sold only for use with qualified monochromated TEMs; these system require extensive site selection and environmental testing. 

Key Features

GIF QuantumERS™

Imaging Filter

Notes

Energy Resolution (eV FWHM) (1)

0.04

2.5mm aperture

Thermal Stability (eV/°C)

0.20

@ 200kV

Isochromaticity (eV P-P max ) (2)

2.0

@ 200kV

Chromatic Distortion (% max over 50 eV) (2)

0.50

@ 200kV

Image Distortion (% max) (2)

0.75


Min./Max. Slit Opening (eV)

2.0 / 100

@ 200kV

Aperture size (mm)

2.5 / 5 / 9


Camera Modes

Normal & Fast


Shutter

1 µs Electrostatic


Energy Field of View (eV)

2000

@ 200kV

Advanced BF/DF Detector

Included


Spectral Rate (/s)

1,000

1x128 binning

   
1.  GIF contribution only.  Does not include TEM source or environmental effects.
2.  Using a 9mm imaging aperture

 

Base Configuration


Hardware

The basic instrument consists of a GIF QuantumERS™ imaging filter with a 2kx2k CCD detector, an entrance flange, computer controlled BF/DF detector, pneumatic aperture assembly, magnetic prism, pre-slit quadrupole assembly, post slit multipole lens assembly, computer controlled slit, and all standard EELS/EFTEM software mentioned in the following paragraph. Where required, a high performance computer workstation is supplied (See Recommended Computer worksheet for details).  For full analytical STEM support, see the model 777 STEMPack product, which includes all the software and hardware required for STEM spectrum imaging.

 

Software

The GIF QuantumERS™ series imaging filters are supported via GMS 1.9, which is exclusive to the GIF Quantum™ series, and the new streamlined AutoFilter® for GMS 1.9 software interface.  GMS software licenses for GIF operation, system tuning, and EFTEM/EELS data acquisition and analysis are included with each GIF Quantum™ imaging filter purchase.  This includes drivers and basic GIF and camera operation software, AutoFilter tuning, EFTEM mapping and quantitative analysis of EELS spectra.  The GIF QuantumERS™ imaging filters comes standard with licenses for basic Gatan EFTEM spectrum image data acquisition and analysis.

 

Optional Components

The following optional components are available for the GIF QuantumERS™ imaging filter:

  •   STEMPack™ spectrum imaging
  •   High-speed spectrum imaging upgrade
  •   2kV high-speed spectrum offset module
  •   DualEELS™ acquisition
  •   EDS Acquisition and Analysis
  •   STEM Diffraction Imaging
  •   Advanced acquisition automation
  •   EFTEM Tomography
  •   Model 806 HAADF STEM Detector.