ORIUS™ SC200W TEM CCD CAMERA
is a 2k x 2k side mount fiber-optical coupled camera that delivers live image viewing at TV rate and large field of view. It is designed for bio-medical imaging applications.
Product Description
The Model 830 SC200W CCD camera is a new member to the ORIUS™ family of high performance CCD cameras. It is based on the same ORIUS SC200 2k x 2k CCD (4M pixels) fiber optically coupled camera but utilizes a single 35mm port of the TEM. This configuration is very beneficial to the life science application as it offers wide field of view of your specimens. The combination of HCR™ (high-contrast resolution) optical technology, and advanced CCD readout electronics has given the SC200W camera the best price-performance value.
It is the fastest 35mm port, digital CCD camera that performs like a TV (30fps). It allows users to capture gain corrected, in-situ events live and allows for the possibility of a direct viewing chamber replacement. It can also be used in conjunction with a high resolution, high sensitivity bottom mount camera to perform general low mag, low dose surveying of the sample. A great combination for true cryo applications.
The power behind the SC200W is it can output high quality (gain normalized) LIVE images that are corrected for cosmetic defects or shading, unlike old analog TV systems that could not.

Kidney recorded at 200x, 120kV Lingual nerve recorded at 13500x, 120kV
Benefits:
-
High performance digital TV CCD camera
-
Low-dose surveying
-
Fastest readout speed (30fps), full CCD
-
Best price-performance value
Features:
· 4 Megapixel CCD sensor (2048x 2048)
· 14-bit dynamic range
· Temperature-regulated Peltier cooling
· IEEE 1394b Firewire interface
· In-situ Observation
· Digital Video Streaming is standard
Specifications
|
Feature |
Description |
|
Operating Voltage |
Up to 120kV |
|
CCD Sensor |
Progressive interline device (2048 x 2048 pixels, 7.4µ each) |
|
CCD Active area |
15mm x 15mm |
|
Anti-blooming |
On-chip |
|
Scintillator |
High-resolution phosphor optimized for kV |
|
Coupling |
Fiber optics 1:1 |
|
Binning |
1x 2x 4x |
|
CCD readout |
Full or sub area |
|
Magnification to film |
1/3 (35mm port) |
|
Readout speed |
30MHz / 5MHz |
|
Dynamic range |
14-bit |
|
Frame rate |
> 30fps @ 4x binning, full CCD area |
|
Peltier cooling |
10°C |
|
Dark current |
< 2 counts / pixel / sec (1x binning) |
|
Readout noise |
< 10 counts |
|
Mounting position |
Single 35mm TEM side port (retractable) |
|
Gain uniformity |
Transparent gain correction and dark subtraction; better than 10% r.m.s. in uncorrected images; better than 0.5% r.m.s. in gain corrected image |
|
Conversion efficiency |
> 6 counts / primary electron (100kV) |
|
Resolution (Nyquist freq.) |
> 0.4 @ 100kV with 1x binning |
|
Full Well |
> 65,000 e- |
|
Non-linearity |
< 2% |
|
Exposure setting |
1msec—30 min |
|
CCD overheat protection |
YES |