WDS

Wavelength dispersive x-ray spectroscopy (WDS) for precise elemental identification and quantification.

Wavelength dispersive x-ray spectroscopy (WDS)
 

Overview

Wavelength dispersive x-ray spectroscopy (WDS) is a remarkable analytical technique that uncovers the elemental composition of a sample by examining the characteristic x-rays it emits. As a powerful companion to energy dispersive x-ray spectroscopy (EDS/EDX), WDS offers enhanced resolution and sensitivity, enabling a precise identification and quantification of elements present in diverse materials.

 

What is WDS?

When an electron beam interacts with a sample, it generates x-rays characteristic of the elements present, providing critical information about material composition. EDS and WDS analyze these characteristic x-rays; however, EDS separates x-rays based on energy, while WDS separates x-rays based on wavelength. WDS offers higher resolution and better accuracy than EDS, which is typically faster and more sensitive to lower-energy x-rays.

 

A WDS system integrated with a scanning electron microscope (SEM) systematically collects and analyzes the emitted x-rays in terms of energy and intensity, enabling precise elemental identification and quantification.

 

The main components of a WDS analysis system include an electron beam source (from the SEM), x-ray collection optics, a diffractor (crystal), and an x-ray detector. The diffraction crystals disperse x-rays by wavelength according to Bragg’s law, which states that x-rays of a specific wavelength are refracted at a particular angle. The rotation of the diffractor is set to direct specific wavelengths, such as the characteristic x-ray of the element of interest, toward the x-ray detector. This detector measures the intensity (flux) of the selected x-rays.

 

WDS resolves EDS peak overlaps, addressing problems that cannot be solved with EDS alone.
Figure 1. WDS resolves EDS peak overlaps, addressing problems that cannot be solved with EDS alone.

Advantages of WDS

WDS enhances peak resolution and signal-to-noise ratio, making it ideal for materials analysis, especially when dealing with complex samples with overlapping x-ray lines or trace elements. Discover how its capability to detect light elements is a game-changer for a wide range of applications.

 

CapabilityAdvantage
Identifies elements in complex samplesOffers better spectral resolution than other x-ray analysis techniques (e.g., EDS), making it ideal for analyzing complex samples that have overlapping x-ray lines (see Figure 1)
Allows the study of trace elementsDelivers a >10x improvement in the minimum detection limit compared to EDS (~0.01 wt.%), enhancing the detection and quantification of trace elements
Examines traditionally problematic light elementsUnique ability to measure low-intensity peaks makes the detection of light elements possible, including beryllium, boron, carbon, nitrogen, and oxygen
Accurately quantifies elementsHighly accurate and precise quantitative analysis enables it to distinguish elements more easily, leading to more reliable and impactful results

Workflow for WDS

Step 1: Sample preparation

Samples must be flat, polished, and mounted on holders or stubs. Non-conductive samples require a coating to eliminate charging effects that could cause image distortion or drift. The best way to accomplish this is to utilize a fully automated argon ion polishing system, like the PECS™ II instrument.

 

A PECS II instrument can produce damage-free surfaces, cross-sections, and deposit coatings to protect or eliminate charging.

 

Specimen Prep YouTube Playlist

Step 2: Data collection

WDS detectors collect emitted x-rays when the microscope beam hits the sample. These x-rays are collected by the WDS detectors, and various crystals in the detectors are used to separate and diffract specific x-ray wavelengths. This provides precise measurement of the intensity of a single x-ray line at a time.

 

The EDAX® APEX™ software makes this conversion possible. WDS can also be combined with related techniques, such as EDS and electron backscatter diffraction (EBSD), to further understand the material’s composition.

Step 3: Analyze

Once the data is collected, APEX software can produce elemental maps, distributions, phase maps, and quantification analyses. The results provided by WDS provide better accuracy and enhance the analysis of complex materials.